PNEUMATIC TUBE TERHADAP DARAH RUTIN DAN LAKTAT DEHIDROGENASE

Authors

  • Liong Boy Kurniawan
  • Asvin Nurulita
  • Uleng Bahrun

DOI:

https://doi.org/10.24293/ijcpml.v21i2.1087

Keywords:

Pneumatic tube, routine blood test, calium, lactate dehydrogenase

Abstract

The transportation of laboratorial samples with pneumatic tube system spends less time than when were handled by courier.
Pneumatic tube system produces minor vibrations on sample due to the velocity changes during transportation and may cause changes
on the blood cells and haemolysis. The aim of this study is to know the effect of sample transportation with pneumatic tube on blood
cells and its effect on hemolysis. A cross sectional study was performed at Dr. Wahidin Sudirohusodo Hospital, Makassar in July 2013.
Routine blood, electrolyte and LDH were tested in 12 out-patients. The researchers collected two (2) samples, for each EDTA tube (routine
blood tests) and serum (electrolyte and LDH). The samples were transported using pneumatic tube and the paired samples were sent
by courier. The result then were analyzed with Paired T-Test. There were no significant difference of routine blood test results between
samples sent by pneumatic tube and courier except RDW. RDW were higher in samples which were sent by pneumatic tube compared
to those brought by the courier (18.72±2.70% vs 17.83±2.36%, p=0.007). The electrolyte levels sent by both methods there were no
significant difference, but the LDH levels were higher in samples sent by pneumatic tube (472.08±100.44 U/L vs 331.25±94.19 U/L,
p=0.000). Based on this study, in common can be concluded that the pneumatic tube system does not effect on the routine blood test
results, except on RDW and does not cause changes due to haemolysis (on electrolyte) except the LDH levels elevates. So based on this
study, it is recommended to send samples for LDH test only by courier.