COMPARISON OF AUTOMATIC STARRSED RS ERYTHROCYTE SEDIMENTATION RATE WITH MANUAL WESTERGREN METHOD

Authors

  • Ninik Sukartini Clinical Pathology Department, Cipto Mangunkusumo Hospital, Faculty of Medicine Universitas Indonesia
  • Siskawati Suparmin Universitas Indonesia

DOI:

https://doi.org/10.24293/ijcpml.v25i3.1302

Keywords:

automated, ESR, Starrsed, Westergren

Abstract

This study aimed to compare the results of ESR measurements using Starrsed RS (30- and 60-minute mode) and manual Westergren method. We also investigate the effect of high leukocytes and haematocrit (Ht) on Starrsed RS ESR measurements. A cross-sectional study using K3EDTA anticoagulated blood samples from 140 subjects were conducted. A total of 21 samples with leukocytes count >50 000/µl were examined. The effect of Ht was studied using data from comparative study subjects.  The 60-minute mode Starrsed RS correlated very strongly [r = 0.98 (95%CI 0.97 to 0.98), p <0.0001], had good agreement [mean bias -0.4 mm (95%CI -1.46 to 0.74)], and gave no significant different results with manual Westergren method. The 30-minute mode Starrsed RS also has a very strong correlation [r = 0.97 (95%CI 0.96 to 0.98), p<0.0001] and good agreement [mean bias-2.9 mm (95%CI -4.23 to -1.58)], but has systematic difference with manual Westergren method. There was significant difference between manual method and Starrsed RS in samples with high leukocytes (p = 0.0208). There were no significant differences of ESR results measured by Starrsed RS dan manual Westergren method in the group of Ht≤35% and >35%. The ESR on the Starrsed RS should still be read at 60 minute. Leukocytes>50 000/µl may affect ESR measurements on Starrsed but no effect of haematocrit was observed.

Downloads

Download data is not yet available.

References

Jou JM, Lewis SM, Briggs C, Lee SH, Salle B, McFadden S. ICSH review of the measurement of the erythrocyte sedimentation rate. Int Jnl Lab Hem 2011; 33: 125-32.

Curvers J, Kooren J, Laan M, Lierop E, Kerkhof D, Scarnhorst V, et al. Evaluation of the Ves-Matic Cube 200 erythrocyte sedimentation method: Comparison with Westergren-based method. Am J ClinPathol 2010;134: 653-60.

RR Mechatronics. A classic gold standard: The Westergren method for ESR measurement [Internet]. Netherlands: RR Mechatronics; c2016. [updated 2016 Feb; cited 2016 Nov 30]. Available from: www.rrmechatronics.com.

Clinical and Laboratory Standards Institute. Procedures for Erythrocyte Sedimentation Rate Test; Approved Standard-Fifth Edition. CLSI document H02-A5. Wayne: The Clinical and Laboratory Standards Institute; 2011.

Rogers R. Erythrocyte sedimentation rate: 30 minute values [Internet]. Netherlands: Medical Laboratory Worlds; c2017 [updated 1993;cited 2016 Nov 30]. Available from: https://rrmechatronics.com/esr/publications/.

Osei-Bimpong A, Burthem J. Supplementary Techniques Including Blood Parasite Diagnosis. In: Bain BJ, Bates I, Laffan MA, Lewis SM, editors. Dacie and Lewis Practical Haematology. 11th ed. China: Elsevier; 2011. p. 102-5.

RR Mechatronics. Starred RS User Manual. Netherland: RR Mechatronics International B.V; 2015.

Manley RW. The effect of room temperature on erythrocyte sedimentation rate and its correction. J Clin Path 1957; 10:354.

Bogdaycioglu N, Yilmaz FM, Sezer S, Oguz E. Comparison of iSED and Ves-Matic Cube 200 erythrocyte sedimentation rate measurements with Westergren method. J Clin Lab Anal 2014;00:1-6.

Linnet K, Boyd JC. Selection and Analytical Evaluation of Methods with Stastitical Techniques. In: Burtis CA, Ashwood ER, Bruns DE, editors. Tietz Textbook of Clinical Chemistry and Molecular Diagnostics. 5th ed. Missouri: Elsevier; 2012. p. 7-48.

Horsti J, Rontu R, Collings A. A comparison between the StaRRsed Auto-Compact erythrocyte sedimentation rate instrument and the Westergren method. J Clin Med Res 2010;2(6):261-5.

Emelike OF, Akpan JE, Obigwe BU, Jeremiah ZA. Comparative study of erythrocytes sedimentation rate (ESR) using trisodium citrate, normal saline, and whole blood in ethylene di amine tetra acetic acid (EDTA). J Appl Sci Environ Manage. 2010; 14 (1): 23-7.

College of Physicias and Surgeons of Alberta. Erythrocyte Sedimentation Rate [Internet]. Alberta: Alberta Laboratory Quality Enhancement Program. c2004 [updated 2000 May; cited 2017 Mar 24]. Available from: https://www.cpsa.ca/wp-content/uploads/2016/08/ESR-Educational-Document-.pdf/.

International Council for Standardizaion in Haematology. ICSH recommendations for measurement of erythrocyte sedimentation rate. J Clin Pathol. 1993;46:198-203.

Clinical and Laboratory Standards Institute. Reference and Selected Procedures for Erythrocyte Sedimentation Rate Test; Approved Standard-Fourth Edition. CLSI document H02-A4. Wayne: The Clinical and Laboratory Standards Institute; 2000.